Autoclaving uses saturated steam under pressure (approximately 15 psi) to bring the temperature in the autoclave to at least 121°C (250°F). Autoclaving can be used to destroy plant-borne bacteria, bacterial spores and viruses.
When disinfecting biohazardous waste, it is recommended that the temperature in the waste reaches at least 115°C for at least 20 minutes. The total processing time required to meet these conditions depends on a number of loading factors (see below); however, when decontaminating waste, an autoclave cycle of at least one hour is recommended.
Please note that biowaste can only be decontaminated using EHS-approved autoclaves, and all decontaminated waste will still be packaged and shipped on-site for destruction as regulated biowaste. Autoclaving these materials simply makes them safer to handle and transport, without affecting the disposal endpoint.
When using an autoclave, the following guidelines should be considered:
Biohazardous materials should not be left in the autoclave overnight in anticipation of autoclaving the next day.
Autoclaves should not be operated by untrained personnel.
Special precautions should be taken to prevent accidental removal of material from the autoclave before sterilizing it or opening both doors simultaneously in a double-door autoclave.
Dry hypochlorite or any other strong oxidizing agent must not be autoclaved with organic materials such as paper, cloth or oil:
WARNING: Oxidizing agents + organic materials + heat = possible explosion
A combination of three factors determines the effectiveness of autoclaving:
temperature
Autoclaves use steam at a pressure of approximately 15 psi to achieve a chamber temperature of at least 121°C. Although the autoclave chamber may reach 121°C, this does not necessarily mean that the interior of the load will reach that temperature.
time
An autoclave cycle time of at least 20 minutes (which does not start immediately after initiating the autoclave cycle) at a chamber temperature of 121°C is generally recommended for sterilizing clean items.
However, the total processing time required to achieve decontamination depends on several loading factors, including:
load vessel (heat transfer characteristics);
The amount of water added to the load;
and the weight of the load.
For increased loads, increased cycle times will be required to ensure effective decontamination.
connect
Steam saturation is critical for maximum heat transfer. Steam must contact all areas of the load. Autoclave bags and other containers should be partially open (or otherwise allow steam to enter) to ensure adequate access. Studies have shown that adding water to the interior of the bag improves the time-temperature profile of the autoclave cycle, thereby increasing the sterilization efficiency of the autoclave.