Correct installation and operation steps of sterilizer

Dec 09, 2016

Correct installation and operation steps (procedure)

Install
Installation is a very important step for the first use of instruments and equipment. **, to ensure that the instrument will not be damaged by incorrect connections or even burn the instrument itself; second, it is a prerequisite to ensure normal use of the instrument. This model (xíng hào) sterilization (Sterilization) has a large power (about 3 kW), so it has higher requirements for the cable line. It must be equipped with a large rated current (Electron flow) (it is recommended to be above 20 A). Three-phase AC power supply to ensure the normal use of other instruments and the safety of the experimental environment to avoid burning the instrument or even causing a fire.
Operation control steps (procedure)
(1) Take out the instrument and clean the cavity of debris. Be sure to refer to the instruction manual carefully before use. If there is a dedicated person to manage it, it is recommended that the user print out a copy of the operating procedures.
(2) Place the instrument in a stable and safe place, turn the manual drain valve to the "closed" position, and add distilled water (usually more than 10 L) into the sterilizer. Add water slowly to avoid adding too much water. This will affect the normal operation of the instrument. When the high water level light is on, add 1 to 2 more L.
(3) The temperature and time can be set according to the specific sterilization items. The specific operations are as follows: Press the "Set" key, and the temperature setting is displayed. You can press the up and down arrow keys to set the value you want. The sterilization time can be set using the same method.
(4) Put the sterilization (fungus) items into the cavity and arrange them neatly. It should be noted that the height of the items should be lower than the cavity to avoid blocking the top exhaust valve and causing danger. High-pressure sterilizers, also known as high-pressure steam sterilizers, can be divided into portable sterilizers and vertical high-pressure sterilizers. A device that uses electric heating wires to heat water to generate steam and maintain a certain pressure. It mainly consists of a sealable barrel, pressure gauge, exhaust valve, safety valve, heating wire, etc. The sterilization method and time of different sterilization materials are different. Special attention should be paid to the sterilization of dressings. The dressings are installed in stainless steel containers. Most users forget to open the ventilation holes of the dressing box, resulting in Because each sterilization cannot reach the color of the test strip, the sterilization fails. Some instruments are sterilized by wrapping them with cloth tape. After sterilization is completed, do not open the wrapping tape, otherwise the infection may occur again.
(5) When sterilizing liquids and culture media, pay attention to the fact that after sterilization is completed, the instrument must be allowed to naturally exhaust and cool down. Do not deflate quickly or cool down rapidly to avoid causing the glass container to collapse due to sudden temperature changes. Burst, thereby threatening the safety of experiment personnel and causing instrument infection. An autoclave is a high-pressure container that uses a pressure higher than normal pressure to raise the boiling point of water to a high temperature above 100°C to sterilize liquids or appliances.
(6) It is best to cover the sterilized items with a layer of cloth so that the sterilized items will not worry about the sterilization effect due to moisture. Try to allow the instrument to cool down naturally. It is prohibited to open the top. Safety valve and exhaust valve to avoid the risk of high-pressure gas injuring people. Generally, the safety valve and exhaust valve above the instrument should be opened once a month to avoid danger caused by corrosion and death.
(7) Tighten the door cover, and turn the manual exhaust (Exhaust) valve under the instrument to the appropriate exhaust position to facilitate the uniform temperature of the chamber gas. At this time, the interlock switch is closed, the interlock light lights up, and the instrument starts to work. (gōng zuò).
(8) When the instrument is working (gōng zuò), the temperature begins to rise. When this instrument leaves the factory, it has an under-voltage beep function. The instrument will keep beeping until the pressure light comes on and the beep stops. High-pressure sterilizer series products are equipment that use pressure saturated steam to quickly and reliably sterilize products. They are suitable for medical and health undertakings, scientific research, agriculture and other units to sterilize medical instruments, dressings, glassware, solution culture media, etc. Sterilization is the ideal equipment.